Motore di ricerca datesheet componenti elettronici
  Italian  ▼
ALLDATASHEETIT.COM

X  

HS4014 Scheda tecnica(PDF) 5 Page - Renesas Technology Corp

Il numero della parte HS4014
Spiegazioni elettronici  High Performance Relative Humidity and Temperature Sensor with Membrane and Digital Output
PDF  27 Pages
Scroll/Zoom Zoom In 100%  Zoom Out
Produttore elettronici  RENESAS [Renesas Technology Corp]
Homepage  http://www.renesas.com
Logo RENESAS - Renesas Technology Corp

HS4014 Scheda tecnica(HTML) 5 Page - Renesas Technology Corp

  HS4014 Datasheet HTML 1Page - Renesas Technology Corp HS4014 Datasheet HTML 2Page - Renesas Technology Corp HS4014 Datasheet HTML 3Page - Renesas Technology Corp HS4014 Datasheet HTML 4Page - Renesas Technology Corp HS4014 Datasheet HTML 5Page - Renesas Technology Corp HS4014 Datasheet HTML 6Page - Renesas Technology Corp HS4014 Datasheet HTML 7Page - Renesas Technology Corp HS4014 Datasheet HTML 8Page - Renesas Technology Corp HS4014 Datasheet HTML 9Page - Renesas Technology Corp Next Button
Zoom Inzoom in Zoom Outzoom out
 5 / 27 page
background image
HS401x Datasheet
R36DS0023EU0103 Rev.1.03
Jun 6, 2022
Page 5
2.
Specifications
2.1
Absolute Maximum Ratings
CAUTION: The absolute maximum ratings are stress ratings only. Stresses greater than those listed below can
cause permanent damage to the device. Functional operation of the HS401x at absolute maximum ratings is not
implied. Exposure to absolute maximum rating conditions might affect device reliability.
Parameter
Conditions
Minimum
Maximum
Unit
Storage Temperature Range
Recommended 0 to 60°C
-40
125
°C
2.2
Recommended Operating Conditions
Important note: The HS401x series sensors are optimized to perform best in the more common temperature and
humidity ranges of 10°C to 50°C and 20% RH to 80% RH, respectively. If operated outside of these conditions
for extended periods, especially at high humidity levels, the sensors may exhibit an offset. In most cases, this
offset is temporary and will gradually disappear once the sensor is returned to normal temperature and humidity
conditions. The amount of the shift and the duration of the offset vary depending on the duration of exposure and
the severity of the relative humidity and temperature conditions.[1] The time needed for the offset to disappear
can also be decreased by using the procedures described in sections 8 and 9.
Parameter
Condition
Minimum
Typical
Maximum
Unit
Operating Supply Voltage
-
1.71
3.3
3.6
V
Sleep Current
Sleep Mode
25°C
-
0.010
0.025
µA
-40 to 125°C
-
-
2.5
Average Current
One RH + temperature
measurement/second
VDD = 3.3V
8-bit resolution
0.27
0.30
0.32
µA
10-bit resolution
0.31
0.34
0.37
12-bit resolution
0.39
0.43
0.47
14-bit resolution
0.55
0.62
0.69
Measurement Time
Humidity and
temperature including
the digital
compensation
8-bit resolution
-
0.64
-
ms
10-bit resolution
-
0.80
-
12-bit resolution
-
1.04
-
14-bit resolution
-
1.70
-
Operating Temperature Range
-
-
-40
-
125
°C
1.
At TA = +25°C, VDD = +1.71V to +3.6V unless otherwise noted.



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27


Scheda tecnica Scarica

Go To PDF Page


Link URL



Lei ha avuto il aiuto da alldatasheet?  [ DONATE ] 

Di alldatasheet   |   Richest di pubblicita   |   contatti   |   Privacy Policy   |   Collegamento alla scheda tecnica    |   scambio Link   |   Ricerca produttore
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com