| Motore di ricerca datesheet componenti elettronici |
|
HMP4 Scheda tecnica(PDF) 2 Page - Vaisala |
|
|
|||||||||||||||||||||||||||||
HMP4 Scheda tecnica(HTML) 2 Page - Vaisala |
|
2 / 3 page ![]() Technical Data Measurement Performance Relative Humidity Sensor HUMICAP R2 Composite Measurement range 0 ... 100 %RH Accuracy at +23 °C (+73.4 °F) 1) ±0.8 %RH (0 ... 90 %RH) T63 response time 15 s Temperature Sensor Pt100 RTD Class F0.1 IEC 60751 Measurement range -70 ... +180 °C (-94 ... +356 °F) Accuracy at +23 °C (+73.4 °F) 1) ±0.1 °C (±0.18 °F) 1) Defined against calibration reference 0.0 0.5 1.0 1.5 2.0 2.5 3.0 Temperature [°C] 3.5 4.0 90 ... 100 % RH 0 ... 90 %RH HMP4 Humidity Measurement Accuracy as a Function of Temperature (Including Non-Linearity and Repeatability) 0 0.1 0.2 0.3 0.4 Temperature [°C] 0.5 HMP4 Temperature Measurement Accuracy over Full Range (Including Non- Linearity and Repeatability) SI Traceable Calibration Uncertainty of relative humidity calibration (k = 2) ±0.5 %RH (0 ... 40 %RH) ±0.8 %RH (40 ... 95 %RH) Uncertainty of temperature calibration (k = 2) ±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F) Operating Environment Operating temperature range for probe body -40 ... +80 °C (-40 ... +176 °F) Operating temperature range for probe head -70 ... +180 °C (-94 ... +356 °F) Operating environment Suitable for outdoor use IP rating IP66 Electromagnetic compatibility Complies with EMC standard EN61326-1, Electrical equipment for measurement, control and laboratory use - EMC requirements - Industrial environment Operational pressure < 100 bar Inputs and Outputs Operating voltage 15 ... 30 VDC Current consumption 10 mA typical 500 mA max. Digital output RS-485, non-isolated Default serial settings 19200 bps N 8 2 Protocols Modbus RTU Output Parameters Relative humidity, temperature, dew point temperature, wet-bulb temperature, absolute humidity, mixing ratio, water concentration, water mass fraction, water vapor pressure, enthalpy Mechanical Specifications Probe fitting M22x1.5 and NPT1/2” fittings included Connector M12/5 Weight 530 g (18.7 oz) Materials Probe AISI316 Probe body AISI316 Cable jacket FEP 136 [5.35] Probe cable 2000 [78.74] M12/5 mm [in] Probe Body Dimensions 183 [7.20] 120 [4.72] Fitting body M22x1.5 or NPT1/2” 41 [1.61] 32 mm 27 mm mm [in] HMP4 Probe Head Dimensions |
|
|
Link URL |
| Lei ha avuto il aiuto da alldatasheet? [ DONATE ] |
Di alldatasheet | Richest di pubblicita | contatti | Privacy Policy | Collegamento alla scheda tecnica | scambio Link | Ricerca produttore All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |